Reset Values of the Status Reporting System    Related Topics  

The table below indicates the effects of various commands upon the status reporting system of the analyzer.

Event

Switching on
supply voltage

Power-On-Status-Clear

DCL,SDC(Device Clear,
Selected Device
Clear)

*RST or SYSTem:PRESet

STATus:PRESet

*CLS

Effect

0

1

 

 

 

 

Clear STB,ESR

 

yes

    

    

    

yes

Clear SRE,ESE

 

yes

    

    

    

 

Clear PPE

 

yes

    

    

    

 

Clear EVENt parts of the registers

 

yes

    

    

    

yes

Clear ENABle parts of all OPERation-and QUESTionable registers,
Fill ENABle parts of all other registers with "1".

 

yes

    

    

     yes

 

Fill PTRansition parts with „1"
Clear NTRansition parts

 

yes

    

    

     yes

 

Clear error queue

yes

yes

    

    

    

yes

Clear output buffer

yes

yes

     yes

     1)

     1)

1)

Clear command processing and input buffer

yes

yes

     yes

    

    

 

 

1) Every command being the first in a command line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR> clears the output buffer.